Three Kinds of SPE Cartridges

Solid phase extraction, as we call it, SPE. Three SPE cartridge types are widely used for the pretreatment when the sample is analyzed. With SPE used for separation, the solution will be generally left behind by gravity or with a vacuum device to speed up the separation, and many of the problems associated with liquid/liquid extraction can be avoided, such as incomplete phase separations and less-than-quantitative recoveries. Hawach provides mixed mode C8/SAX SPE cartridges, mixed mode C8/SCX SPE cartridges, polymeric MCX SPE cartridges, and other SPE cartridges of your choice.

The scientists in the lab can also save lots of expensive, breakable specialty glassware, and they don’t need to dispose of large quantities of organic solvents.

Three kinds of SPE cartridges

According to a similar dissolution mechanism, there are three SPE cartridge types according to the packing materials in the SPE cartridges, namely, reverse phase extraction, normal phase extraction, and ion exchange extraction.

Cartridge TypeSorbent PolarityRetention MechanismBest For
Normal PhasePolarPolar interactionsPolar compounds in non-polar matrices
Reversed PhaseNon-polarHydrophobic interactionsNon-polar/moderate compounds in water
Ion-ExchangeChargedIonic interactionsCharged analytes (acids/bases)
  1. 1. Normal phase SPE cartridges:

    The adsorbents used in normal phase SPE are all polar, such as bonded silica gel, alumina, silica-silver, Florisil (2MgO·3SiO2), etc. to extract polar substances. Such as Normal Phase Silica SPE Cartridge.

    Sorbent Type:

    • Polar materials (e.g., unmodified silica, diol, amino)

    Mechanism:

    • Retains analytes via polar interactions (e.g., hydrogen bonding, dipole-dipole)
    • Non-polar or less polar substances elute first

    Typical Applications:

    • Extraction of polar compounds from non-polar matrices (e.g., environmental samples, oils)
    • Separation of structural isomers and polar metabolites

    Elution Strategy:

    • Elution with more polar solvents (e.g., methanol, water)

    2. Reversed-phase SPE cartridges:

    The adsorbent used is usually non-polar or weakly polar, such as C8, C18, phenyl column, etc. The target compound to be extracted is usually medium to non-polar. Such as Reversed Phase C18 SPE Cartridges.

    Sorbent Type:

    • Non-polar or moderately polar bonded silica (e.g., C18, C8, phenyl)

    Mechanism:

    • Retains analytes through hydrophobic interactions
    • More polar components elute first

    Typical Applications:

    • Isolation of non-polar to moderately polar compounds (e.g., drugs, hormones, pesticides) from aqueous samples
    • Cleanup of biological samples (e.g., plasma, urine)

    Elution Strategy:

    • Elution with organic solvents (e.g., methanol, acetonitrile) after washing with water or buffers

    3. Ion-exchange SPE cartridges:

    The adsorbent used for the compound ion exchange type SPE is a charged ion exchange resin such as LC-NH2 (containing aliphatic quaternary propyl bonded silica gel), LC-SCX (containing aliphatic sulfonic acid bonded silica gel), etc. The target compound extracted is a charged compound. Such as Ion-Exchange SAX SPE Cartridge.

    Sorbent Type:

    • Charged functional groups (e.g., sulfonic acid for cation exchange, quaternary amine for anion exchange)

    Mechanism:

    • Electrostatic interaction between the sorbent and oppositely charged analytes
    • Specific to the analyte’s ionic charge

    Subtypes:

    • Cation Exchange: For positively charged compounds (e.g., basic drugs)
    • Anion Exchange: For negatively charged compounds (e.g., acidic metabolites)

    Typical Applications:

    • Extraction and cleanup of charged analytes (e.g., amino acids, peptides, organic acids)
    • Removal of salts or interfering ions from samples

    Elution Strategy:

    • Elution with solutions that disrupt ionic interactions, often by changing pH or ionic strength
SPE Cartridges C18
Diol SPE Cartridges.png

Principle of SPE Cartridge

Three SPE cartridge types provided by Hawach Scientific are a separation technique combining selective retention and selective eluting. It also works as a sample pretreatment device for extraction, separation, and concentration. Currently, it is widely applied to the analysis of various foods, environmental and biological samples, and so on.

When the liquid sample solvent passes through the adsorbents, it retains the tested substances, then selects the appropriate strength solvent to flush away the impurities, and then quickly elutes the tested substances with a small amount of solvent, so as to achieve the purpose of rapid separation, purification, and concentration.

SPE(Solid-phase Extraction), is an increasingly popularized technology of sample pretreatment, which is established on the basis of traditional LLE(Liquid-liquid Extraction). Combined with the substance interaction mechanism that like dissolves like, and basic knowledge of stationary phases with currently widely applied HPLC(High-Performance Liquid Chromatography) and GC(Gas Chromatography), the skill has been gradually developed.
Besides, SPE features a small number of organic solvents, convenience, safety, and high efficiency.

SPE Technique Typical Applications

  • Biological fluid analysis;
  • Analysis and monitoring of drinking water, groundwater, and sewage;
  • Volatile oils, plant tissues, fruits, vegetables, and grains;
  • Liquid drug samples, soils, and sediments;
  • Meat, fish, and other animal tissues;
  • Tablets and other solid drugs.

Usage of SPE Cartridge

In daily application, the SPE Cartridge can selectively adsorb impurities and allow the tested substances to flow out, or simultaneously adsorb impurities and the tested substances, and then selectively elute the tested substances with the appropriate solvent.

In reaction, the compound is adsorbed by reverse SPE or forward SPE and retained on the SPE packing, and the bonded phase has strong polar adsorption in the non-polar organic solvent and weak anion exchange retention.

SPEs are mostly utilized to treat liquid samples extracting, concentrating, and purifying the semi-volatile and non-volatile compounds. If solid samples are provided, they have to be processed into liquids ahead. Nowadays, domestic applications mainly focus on the analysis of polycyclic aromatic hydrocarbons(PAHs) and polychlorinated biphenyls(PCBs) for watery conditions, pesticide and herbicide residues in fruits, vegetables and food, antibiotic analysis, and clinical drug analysis.

SPE is an advanced way to prepare liquid samples and extract semivolatile or nonvolatile analytes. Scientists often use it with solids that are pre-extracted into solvents as well. Compared with liquid/liquid extraction, SPE is more efficient and easy to perform.

Available in a wide variety of chemistry, adsorbents, and sizes, Hawach SPE cartridges are excellent tools for sample extracting, concentrating, and cleanup. With the consist of a medical PP empty tube, a medical PE sieve plate, and a fixed adsorbent, you can find Hawach SPE Cartridges in different types of silica gel, polymer, adsorption, and the special cartridge according to matrix type.

SPE is a physical extraction process consisting of a liquid phase and a solid phase. During the extraction process, the adsorption of the solid relative analyte is greater than that of the sample base.

When the sample passes through the solid phase cartridge, the analyte is adsorbed on the surface of the solid filler, and the other sample components pass through the cartridge analyte and can be eluted with a suitable solvent to achieve separation and purification.