How to Ensure Good LC Column Performance and Life?
How to ensure good column lifetime during HPLC using a liquid chromatography column, and how to ensure good HPLC column performance? Just pay attention to the following points.
- Carefully read the instruction manual of the column and use and maintain the column as described in the instruction manual. Always use a packed column with good packing when purchasing the column.
- Minimize pressure fluctuations and avoid mechanical and thermal shocks. Use guard columns and inline filters as much as possible during column analysis
- Often flush the column with a strong solvent, fully filter the sample and mobile phase, and try to avoid impurity particles and strong solvent residue.
- Try to use a stable stationary phase. When operating at a medium pH of pH 6-8, use an organic buffer solution. The column temperature is preferably less than 40 degrees. If it is a silica gel matrix column, Keeping the mobile phase’s pH range between 2 and 8 is the safest.
- When the mobile phase contains a buffer solution, care should be taken to use 95:5 water and organic solvent transition, the organic solvent cannot be less than 5%. Do not use pure water to rinse the column of the hydrophobic silica matrix. If the mobile phase must be pure water, please buy reversed-phase column with polar sealing or a polar setting.
- When the column is not to be stored, rinse off the salt and buffer and store with pure organic mobile phase, preferably acetonitrile.